USING THE KICKSTAND FOR DIALYSIS SCALE
The kickstand is intended for weighing patients on a scale .
Note:
• The kickstand assembly is configured for an X-frame cot retention system only .
• The kickstand (p/n 6085-002-000) is not compatible with the optional base storage net (p/n 6500-160-000) .
WARNING
• Stryker recommends a two person operation when using the kickstand .
• Make sure that the patient weight is centered on the cot before using the kickstand .
• Engage the kickstand with your foot only .
• Lower cot height prior to engaging kickstand for increased stability .
• Make sure that the kickstand remains in the retracted position and does not engage during transport .
• Do not use the kickstand as a brake .
• Do not engage kickstand on a sloped surface .
To use the kickstand:
1 .
Operator 1 engages the kickstand with their foot as shown in Figure 51 .
2 .
Operator 2 lifts the foot end of the cot at a height sufficient to actuate the kickstand .
3 .
Both operators must make sure that the kickstand is in the forward locked position as shown in Figure 52 .
Figure 51
To release the kickstand:
1 .
Operator 1 lifts the foot end of the cot until both wheels
are off of the floor .
2 .
Operator 2 rolls the cot slightly forward to make sure
that the kickstand retracts on its own as shown in
Figure 53 .
www.stryker.com
Optional Accessories
6086-009-005 REV A
Figure 52
Figure 53
Return To Table of Contents
English
1-45