Combination Therapy - Pari Pep S System Instrucciones De Uso

Ocultar thumbs Ver también para PEP S System:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 162
Info: Make sure that the blue inhalation valve washer is
seated correctly.

Combination therapy

• Assemble the nebuliser (without mouthpiece) as de-
scribed in the accompanying instructions for use.
• Attach the PEP S connecting element to the nebuliser.
Info: The PEP S inhalation valve and the mouthpiece
with exhalation valve are not needed for combination
therapy.
• Attach the mouthpiece (without exhalation valve) to
the connecting element.
Connecting the pressure gauge
• Open the cap on the PEP S connecting element.
 CAUTION! Risk of infection due to contamin-
ation! If the connection tubing is connected directly to
the PARI PEP S System, germs may build up in the
tubing. Since the connection tubing cannot be disin-
fected, use the pressure gauge exclusively with the
extra pressure gauge connector line and the tubing
adapter, in order to avoid a risk of infection.
Push the pressure gauge connector line onto the pres-
sure gauge attachment point on the PEP S connecting
element.
• Insert the corresponding end of the connection tubing
in the pressure gauge.
• Connect the connection tubing and the pressure gauge connector line using the tubing ad-
apter.
13
Tabla de contenido
loading

Este manual también es adecuado para:

018018g4200

Tabla de contenido