2. Check to see if the JET Air Polishing Insert is secured
inside the Jet-Mate™ handpiece. Tap-On
disabled when there is no JET Air Polishing Insert in the
handpiece.
Symptom:
System will not operate: No Power ON indicator
1.
Check that the Main Power Switch is in the ON (l)
position, and that the detachable Power Cord is fully
seated in the receptacle on back of System.
2.
Check that the system's power cord plug is fully seated
in an approved AC wall outlet.
3.
Check that the wall outlet is functional.
Symptom:
System will not operate: Power ON Indicator is
illuminated
1.
If the office has more than one Tap-On
test each to ensure that the proper Tap-On
is being used. With a Jet-Mate™ handpiece and JET
Air Polishing Insert installed, depress the Tap-On
Pedal to the first position. The system should dispense
water. If none of the Tap-On
system, continue to the next step.
2.
Resynchronize one Tap-On
(see Section 7.10 Tap-On
Symptom:
System operates: No water flow to air polishing
insert tip or handpiece overheats
1.
Assure that handpiece lavage control is properly
adjusted.
2. Check for clogged JET Air Polishing Insert. Replace insert
if necessary.
3.
Check that dental office water supply valves are open.
4.
If the system is connected to DualSelect Dispensing
System, check that fluid level in the selected bottle is
sufficient. Make sure valves are open when using
external water source.
5.
Check that the water line filter is clean. Replace
filter if needed.
Symptom:
System operates: Purge Mode will not function –
icon flashing
1.
Check that there is no JET Air Polishing Insert in the
handpiece.
2.
Check that Jet-Mate™ handpiece is properly installed to
the cable assembly.
Symptom:
System operates: Service Indicator blinking
Fast Blinking (3 blinks per second)
– Indicates improper set-up
1. If JET Air Polishing Insert is in the Jet-Mate™
handpiece, remove. Verify the Jet-Mate™ handpiece
is properly seated and hold the foot control for 2
seconds. If blinking stops, the system is ready for
use. If blinking remains, continue to the next step.
2. Attach a NEW Jet-Mate™ handpiece and hold Tap-
On
Foot Pedal for 2 seconds. If blinking stops, the
TM
Technology is
Foot Pedal,
TM
Foot Pedal
TM
Foot
TM
Foot Pedals operate the
TM
Foot Pedal to the system
TM
Foot Pedal Synchronization).
TM
system is ready for use. Discard the old handpiece
or return if within warranty. If blinking remains,
continue to the next step.
3. Install and fully seat a JET Air Polishing Insert into
Jet-Mate™ handpiece. Hold Tap-On
2 seconds. If blinking stops, the system is ready
for use. If blinking remains, continue to the next
step.
4. Install and fully seat a NEW JET Air Polishing Insert
in Jet-Mate™ handpiece and hold Tap-On
Pedal for 2 seconds. If blinking stops, system is
ready for use. Discard old JET Air Polishing Insert
or return if within warranty. If blinking remains, refer
to Section 11.2 Technical Support and Repairs to
have unit serviced as soon as possible.
Slow Blinking (1 blink per second)
-- The system is not operating within factory
specifications.
1. Remove JET Air Polishing Insert.
2. Turn Main Power Switch OFF, (O) position. Wait five
seconds. Turn unit ON, (I) position.
3. Operate Purge function.
4. If service indicator still blinks, refer to section 11.2
Technical Support and Repairs to have unit
serviced as soon as possible.
Symptom:
System operates: Service Indicator illuminated
1.
Ensure that the base unit has adequate ventilation and
is not near a heat source (i.e. radiator, heat lamp,
sunlight or other heat producing operatory equipment).
2.
Turn Main Power Switch to the OFF (O) position.
Allow system to cool for 10 minutes and turn system
to the ON (I) position. Verify light is not illuminated.
3.
If light is still illuminated, refer to section 11.2
Technical Support and Repairs to have unit serviced as
soon as possible.
Symptom:
System operates: Air Polishing Insert nozzle
blocks repeatedly
1.
Powder is contaminated (lumpy). Discard powder.
2.
Air Supply Line Air Filter is contaminated. Refer to
Section 10.4 Air Supply Line Air Filter Maintenance.
3.
Dental office air source should be serviced to eliminate
the source of the contamination.
Symptom:
System operates: No bleed air
1. Blocked JET Air Polishing insert nozzle. Clean nozzle
using supplied tool.
2. Blocked air bleed "duckbill" air filter. Refer to Section
11.2 Technical Support and Repairs to have unit
serviced as soon as possible.
Symptom:
System operates: No or poor cleaning action
1. Very low powder level or empty powder bowl. Fill
Powder bowl.
2. Blocked JET Air Polishing insert nozzle. Clean nozzle
20
Foot Pedal for
TM
Foot
TM