08.2 Û CLEANING OF THE DEVICE BODY
PREPARATION
- Make sure that all the following accessories
have been removed/disconnected from the
device body (see Chapter 07):
• electrical power supply cable
• foot pedal
• handpiece
• bottle
MATERIALS NECESSARY
- Clean, soft cloths with low
fiber-release
- Detergent solution (pH 6-9) and, if need be,
non-aggressive disinfectant solution with
neutral pH (pH7)
CLEANING AND DISINFECTION METHOD - Device body
08 Û CLEANING AND STERILIZATION
WARNING: Do not sterilize the device
body: it could stop working and cause
harm to person and/or damages to things.
WARNING: Always turn the device off
via the switch and disconnect it from the
electrical network before performing the
cleaning and disinfection interventions.
WARNING: The device is not protected
against penetration of liquids. Do not spray
liquids directly onto the surface of the
device.
CAUTION: We recommend that you use
water-based disinfectant solutions with
neutral pH (pH7). Alcohol-based disinfectant
solutions and hydrogen peroxide are contra-
indicated, because they can fade the color
and/or damage the plastic materials. This
also holds true for chemical products such
as acetone and alcohol.
Clean the surface of the device and of
1
the handpiece cord with a clean, soft
cloth with low fiber-release, dampened
with a detergent solution (pH 6-9) and, if
need be, disinfect with a non-aggressive
disinfectant solution with neutral pH (pH7),
following the instructions provided by the
manufacturer of the solution;
Dry the device body and the handpiece cord
2
with a clean, non-abrasive cloth with low
fiber-release.
EN
25