Safety and performance
Declaration: Electromagnetic emissions and immunity for equipment and systems that are not life-supporting and are specified for use
only in a shielded location.
Declaration: Electromagnetic immunity
The device is intended for use in the electromagnetic environment specified below. The customer or the user of the device should ensure that it is used in such an environment.
Immunity test
IEC 60601 Test Level
Compliance Level
Conducted RF
3 Vrms
N/A
IEC 61000-4-6
150kHz to 80MHz
Radiated RF
3 V/m
N/A
IEC 61000-4-3
80 MHz to 2.5 GHz
Electrostatic discharge (ESD)
Contact: ±8 kV
Contact: ±8 kV
IEC 61000-4-2
Air: ±2 kV, ±4 kV, ±8 kV, ±15 kV
Air: ±2 kV, ±4 kV, ±8 kV, ±15 kV
Electrical fast transient/burst
2 kV for power supply lines
N/A
IEC 61000-4-8
1 kV for input/output lines
Surge IEC 61000-4-5
1 kV differential mode
line(s) to line(s):
2 kV common mode
+/- 1kV
Voltage dips, short
-5% UT (95% dip in UT) for 0.5
N/A
interruptions and voltage
cycles, -40% UT (60% dip in UT)
variations on power supply
for 5 cycles, -70% UT (30% dip in
input lines IEC 61000-4-11
UT) for 25 cycles, -5% UT (95%
dip in UT) for 5 sec
Power frequency (50/60Hz)
30 A/m
30 A/m
magnetic field IEC 61000-4-8
50 Hz or 60 Hz
50 Hz or 60 Hz
Electromagnetic environment – guidance
N/A
Portable and mobile RF communications equipment should be used no
closer to any part of the equipment or system including cables than the
recommended separation distance calculated from the equation applicable
to the frequency of the transmitter. Interference may occur in the vicinity of
equipment marked with the following symbol:
Floors should be wood, concrete, or ceramic tile. If floors are covered
with synthetic material the relative humidity should be at least 30%.
The main power quality should be of the kind used in a typical commercial or
hospital environment
The main power quality should be of the kind used in a typical commercial or
hospital environment
The main power quality should be of the kind used in a typical commercial
or hospital environment. If the user of the equipment or system requires
continued operation during power main interruptions, it is recommended
that the equipment or system be powered from an uninterruptible power
supply or a battery.
The power frequency magnetic field should be at levels characteristic
of a typical location in a typical commercial or hospital environment.
Federal Communication Commission Interference Statement
This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) This device may not cause
harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired oper-
ation.
FCC Caution: Changes or modifications not expressly approved by the party responsible for compliance could void the user's authority
to operate the equipment
This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the FCC Rules.
These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment gen-
erates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful
interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this
equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and
on, the user is encouraged to try to correct the interference by one or more of the following measures:
- Reorient or relocate the receiving antenna.
- Increase the separation between the equipment and receiver.
- Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
- Consult the dealer or an experienced radio/TV technician for help.
IMPORTANT NOTE:
Radiation Exposure Statement: This equipment complies with FCC radiation exposure limits set forth for an uncontrolled environment.
This equipment should be installed and operated with a minimum distance of 20 cm between the radiator & your body.
This transmitter must not be co-located or operating in conjunction with any other antenna or transmitter.
Country Code selection feature to be disabled for products marketed to the US/CANADA.
Contains FCC ID: 2AAGMGM02SA
MODE
FREQUENCY BAND (MHz)
MAXIMUM OUTPUT POWER (dBm)
LTE
1710 - 1785
23 dBm
LTE
880 - 915
23 dBm
832 - 862
23 dBm
LTE
LTE
703 - 748
23 dBm