Technical Specifications
Product Name: Withings Body Pro 2
Model: WBS14
Battery Life: 10 months typical use
Power Source: Four (4) alkaline cells - AAA
Heart Rate Measurement Range: 30 bpm to 230 bpm
Display: Color 2.8 inch LCD screen
Weight Withings Body Pro 2: 2.2kg
Operation time: 30 seconds
Parts in contact with the skin: Glass plate of the scale
Weight Units: Kilograms (kg), Pounds (lb), Stone (st lb)
Weight range: 5kg - 200kg I 11lb - 440lb I 1st - 31 st
Transport and storage conditions:
-10°C (14°F) — 50°C (122°F)
10% — 90%
86kPa — 106kPa
Max altitude: 2000 m
Operation conditions:
5 °C (41 °F) — 35 °C (95 °F)
20 % — 80 %
86kPa — 106kPa
Max altitude: 2000 m
IP Protection level:
IP22
Mode of operation: Continuous Operation
Minimum product life: 3 years
Please be aware that the stated 3-year product life of Withings scales
represents a minimum lifespan under standard tested conditions.
Withings products are designed to last: our scales are manufactured
with care, and are provided with continual over-the-air firmware
updates to avoid obsolescence. This is why with proper care and usage,
our health products are likely to serve you reliably for much longer than
the minimum product life. In fact, we're proud to share that 50% of our
users are still actively using their Withings scales after 7 years.
Wireless information
Wireless Specifications
Mode
Frequency Band (MHz)
Maximum Output Power (dBm)
BT LE
2402-2480
5 dBm
WLAN
2412-2462
16 dBm
The wireless communication of the device is supported by BLE,
WiFi and cellular communication. This communication is estab-
lished between the device and the Withings application. The com-
munication between the device and the Withings application is
encrypted through an exchange of a paired key.
The communication latency between the device and the
Withings application takes less than 10 seconds when the device
and the smartphone are less than 16 feet (5 meters) apart.
The communication between the device and the Withings applica-
tion is not modified with sources of interference located within 5
meters. Wireless coexistence has been tested in accordance with
the following standards:
- ANSI C63.27:2017 and,
- AAMI TIR69:2017/(R2020)
Electromagnetic disturbances have been tested in accordance
with the standard ANSI AAMI IEC 60601-1-2:2014.
Safety and performance
RF Statement
Medical electrical equipment needs special precautions regarding EMC and needs to be installed and put into service according to the
EMC information provided in the accompanying documents.
Portable and mobile RF communications equipment can affect medical electrical equipment
Guidance and manufacturer's declaration-electromagnetic emissions
Withings Body Pro 2 is intended for use in the electromagnetic environment specified below. The customer or the user of the Withings Body Pro 2 should ensure that it is used in such an
environment.
Emissions Test
Compliance
Electromagnetic environment – guidance
CE emissions CISPR11
Group 1
The device uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference
in nearby electronic equipment.
RE emissions CISPR11
Class B
Harmonic emissions
N/A
IEC 61000-3-2
The device is suitable for use in all establishments, including domestic establishments and those directly connected to the public low-voltage
power supply network that supplies buildings used for domestic purpose.
Voltage fluctuations/Flicker
N/A
emissions IEC 61000-3-3