6 Mechanical conversion

6.1 Preparing the unit for conversion

16
Converting Instructions "ORTHOPHOS XG 3D
6
Mechanical conversion
Mechanical conversion
6.1
Preparing the unit for conversion
Preparing the unit for conversion
1. Switch the unit on and move it to a comfortable working position
(board DX1 must be easily accessible).
2. Switch the unit off again.
WARNING
Potentially lethal shock hazard!
It is essential to switch the unit off and to wait at least another 4 minutes
before removing a cover.
3. Remove the Pan sensor.
Keep this in a safe place for reassembly.
6.2

Removing the cover parts

Removing the cover parts
When removing covers, always remember that direct sunlight or bright
room lighting can cause system malfunctions due to activated light
barriers. Therefore: avoid direct sunlight and bright room lighting above
the unit!