How many result records can the VS2 hold?
The VS2 stores the results of the last 5000 rotors run. Thereafter, the VS2 replaces the oldest
record with the most recent.
What maintenance does the VS2 require?
The only required maintenance is cleaning the air filter on the back of the analyzer.
See
"Cleaning the Air Filter" on page
2.5.2
Analyzer Setup Questions
Does the VS2 require calibration?
No. Unlike any other system, the VS2 reagents are factory-calibrated. Determined by Abaxis
quality control methods within the factory, calibration factors provide mathematical adjust-
ments interpreted by the analyzer from the rotor's barcode to correct for lot to lot reagent vari-
ations.
Does the VS2 need surge protection or an uninterruptable power supply?
Abaxis recommends plugging the VS2 into a surge protector designed for use with a computer.
Abaxis also recommends using an uninterruptable power supply (UPS) in areas prone to elec-
trical surges or power outages.
Can I connect the VS2 to an external computer?
Yes. Sample results and status reports can be transmitted to a computer (through a USB or net-
work port). You can use popular terminal emulation software to collect the data from the VS2.
The data is delivered in raw ASCII format, and can easily be converted to text for use in other
applications.
For information about optional Ethernet connectivity capabilities, or ASTM or XML connec-
tions, contact Abaxis Technical Support, 24 hours a day, 7 days a week, at 1-800-822-2947.
The VS2 also interfaces with many commercial veterinary data management software pack-
ages. For information on compatible packages, check with your software vendor, or contact
Abaxis Technical Support.