VACUUM SETTING

From the SETUP, select VACUUM SETTINGS, default setting is displayed:
Allows to modify the default vacuum time and the default time of check.
Each value can be modified, within the values shown in parentheses.
NOTE: press DEFAULT to restore default values:
− Vacuum time
− Time of check
− Vacuum rising

OPTIONS

Select OPTIONS, insert the OPTION code:43210791 then press ANALYZER, the following
screen will be displayed:
Choose ON to enable gas analyzer or OFF to disable it then press ENTER.
INSTRUCTIONS MANUAL
SETUP
VACUUM SETTINGS
Vacuum time
(Min 2, Max 120 min)
Time of check
(Min 1, Max 120 min)
Vacuum rising
(min 100, max 900 mbar
25 min
2 min
0,1 mbar
ENABLE ANALYZER
OFF
ON
-51-
DD/MM/YY XX:XX PM
ESC
25
2
DEFAULT
100
ENTER
DD/MM/YY XX:XX PM
ENTER