1Manual Gauging Of Decoder (U111Dc) - Silca Tri-Code Manual De Uso

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Operating manual - English
- At the end of the operation the following message will appear (e.g.):
Press [ENTER] to save the data automatically adjusted.
Attention: if the [STOP] key is pressed the new calibrations will be lost and only the previous calibrations will
remain valid.
- if this is not so, the display will show an error message:
In this case the procedure MUST be repeated, carefully following all the steps described; in particular
check the positioning of the clamp (Ch. 4.4, page 18).
Wrong positioning can throw out the calibration.
Note: the calibrations will be accepted only if the variations remain within a range of -40 or +40 hundredths of a mm
(-0.0157" / +0.0157").
5.6.3.1
MANUAL GAUGING OF DECODER (U111DC)
Note: this operation has already been carried out during final machine testing. It should be done only in
exceptional circumstances, such as: replacing the dial or decoder spindle, or if the operator makes
a mistake when replacing the decoder U111DC).
- Raise the safety shield, loosen the grub screw (K) (fig. 37) and remove the U111DC
decoder (Ch. 7.12, page 80).
- Loosen grub screw (K1) (fig. 38).
Fig. 37
- Fit the decoder U111DC and use a screwdriver to turn the decoder height regulation cam screw (K2)
so that when the decoder is fitted it can be positioned as high as possible. Tighten the grub screw
(K) to secure the decoder (fig. 39).
Fig. 39
U 1 11 D C :
X = - 0 1
S a v e d a t a ?
N o = S T O P
Ye s = E N T E R
E x c e e d e d s e t t i n g
t o l e r a n c e l i m i t !
X = + 4 4
Y = + 0 1
S e e o p e r a t i n g m a n u a l .
K
F1
U111DC
K
Copyright Silca 2006
Y = - 1 8
F2
O
K2
TRI-CODE H.S.
U111DC
K1
Fig. 38