Lan Communication; Error Code - Agilent Technologies IPCMini Manual De Usuario

Tabla de contenido

Publicidad

Technical Information

LAN communication

The IPCMini with the LAN option is able to encapsulated the "windows serial
protocol" within a TCP/IP framework. You need to configure the IP address and the
subnet mask from the display (Menu → Communication—LAN). To be sure that the
configuration is done correctly you can ping the controller at the IP address
selected.
After the configuration you can communicate with the controller with TCP
at port 23.

Error Code

The error code is displayed in decimal format. Binary representation of this number
provides indications regarding the type of error that has occurred.
The error table is shown below:
Table 12 Error Code
Error code in decimal format
0
4
32
64
128
200
Binary
0000000000000
0000000000100
0000000100000
0000001000000
0000010000000
Agilent IPCMini Ion Pump Controller User Manual
Description
No error
Over Temperature
Interlock cable
ShortCircuit
Protect

Publicidad

Tabla de contenido
loading

Tabla de contenido