Instructions for Use

Startup and Operation of the IPCMini Controller

To switch on the High Voltage (HV), the interlock of the HV cable (cable interlock)
NOTE
must be closed (connector inserted).
Proceed as follows to power the controller and apply voltage to the high voltage
(HV) connectors:
Connect the power cord to the rear panel.
Switch off the controller.
Connect the HV cable and the interlock of the HV cable in the related
connectors on the rear panel.
Verify that the LOCAL mode is enabled, by pressing "MENU" then "MODE"
push button in sequence.
Keep the HV ON/OFF push button pressed for at least 3 seconds.