UPS Universal Power Supply – Model # 200-006R Instructions For Use
For use only with
Microline Cautery Instruments and Microline PowerPack™ Dual Control Footswitch
Note: Additional IFUs may be obtained free of charge by contacting Microline Customer service, or by visiting
https://www.microlinesurgical.com/resources/ifus/.
Device Description
The Model # 200-006R Universal Power Supply (UPS) is a non-sterile, reusable, AC powered unit intended to be
used only with Microline Cautery Instruments and the PowerPack Dual Control Footswitch. The UPS is intended
to be connected to AC power via a grounded hospital grade power cord and features an on/off switch and green
power-on LED indicator.
Depending on the device used, up to three operating heat ranges are available from the UPS with corresponding
tones indicating the power level of the instruments heating element. A low frequency tone indicates the use of low
heat. An interrupted low tone indicates the use of variable heat, and a higher frequency continuous tone indicates
the use of high heat levels.
If the instrument does not require the use of the PowerPack Dual Control Footswitch, the UPS will supply power at
the variable heat level only, unless a multi-heat range device is used. The UPS may be placed on a flat non-sterile
surface adjacent to the sterile field or may be suspended from a nearby IV pole. If suspended from an IV pole, the
UPS may be rotated to face the user.
Intended Use
1. The UPS is intended for use only with Microline Cautery Instruments and the PowerPack Dual Control
Footswitch for the simultaneous cutting and cauterization of soft tissue during surgery.
Operating Sequence
1. Connect grounded, hospital grade power cord to the UPS AC connector receptacle.
2. Plug the power cord into a grounded hospital grade receptacle.
3. Align the key portion of the Microline Cautery Instrument connector with the UPS connector receptacle and insert.
4. If the PowerPack Dual Control Footswitch is to be used, plug the Footswitch connector into the Footswitch
connector receptacle of the UPS.
5. Turn on the UPS.
6. Refer to the appropriate Microline Cautery Instrument and PowerPack Dual Control Footswitch (if applicable)
Instructions For Use for device specific use instructions and pre-check requirements.
Microline Surgical, Inc. 迈克兰器械有限公司.
50 Dunham Road, Suite 1500
Beverly, MA 01915 U.S.A.
Email:
info@microlinesurgical.com
ENGLISH
09-39-00119-00-ART rev H
1
Tabla de contenido
loading

Resumen de contenidos para Microline Surgical 200-006R

  • Página 53 Descripción del dispositivo El Adaptador de corriente (UPS) Modelo 200-006R es una unidad no estéril, reutilizable que funciona con CA y está prevista para utilizarla solo con los Instrumentos cauterizadores y el Conmutador de pedal de control dual PowerPack de Microline. El UPS está previsto para ser conectado a corriente alterna (CA) mediante un cable de alimentación con puesta a tierra de grado hospitalario, y cuenta con un interruptor de encendido/apagado (On/Off)
  • Página 54 Si no se puede resolver el problema o no se restablece la electricidad después de varios minutos de enfriamiento, solicitar instrucciones a Microline Surgical con respecto a la reparación o el reemplazo del dispositivo.
  • Página 55 • Los equipos de comunicaciones de RF portátiles y móviles pueden afectar a la fuente de alimentación universal. Consulte la información de EMC proporcionada en el Manual de servicio. • El 200-006R no está pensado para ser utilizado en contacto directo con el corazón, y actualmente pertenece a la clasificación «BF».
  • Página 56: Protección Del Medio Ambiente

    Microline Surgical no se responsabiliza por ningún tipo de daño incidental o emergente. La alteración del instrumento anulará esta garantía. El abuso o uso negligente de un instrumento de Microline Surgical conllevará la anulación de esta garantía. Definición de los símbolos...

Tabla de contenido