Operating Instructions BA ULTIMATE
65 56 083 D3560.201.10.02.02
NOTICE! Using sprays from other manufacturers can reduce the
Using tbe BA Ultimate spray
product's service life. Only use BA Ultimate Spray.
IMPORTANT: Use a soft, clean, and disinfected brush for cleaning.
Use soft brush
✔ Wear appropriate protective clothing.
Procedure _ Turbine _ without Alpro
✔ All disinfectants must be approved in your country and have proven
bactericidal, fungicidal and virucidal properties. Only use
disinfectants with no protein-fixing properties.
1. Brush the turbine under running water (< 38 °C, < 100 °F, at least
drinking water quality) and good lighting (min. 500 lux) and color
rendering index (min. 80 Ra) until no more dirt can be seen, for at
least 10 seconds.
2. Flush the drive channels with spray.
3. Clean and disinfect spray channels with suitable agents and
adapters according to manufacturer's instructions.
4. Blow spray channels with suitable adapters out with 2.5 - 3 bar until
no more moisture displaces, but at least 10 seconds.
7 Conditioning