00Dmean; 00Dsemi; 00Dcycl; 00Dstop - marposs TESTAR Quick read M Manual De Instalacion Y Uso

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 127

00DMEAN

Request from Host
Response from QuickRead(s)
Select Dynamic Type Measure of (Maximum + Minimum)/2.
All QuickRead(s) check the possibility and subsequently select dynamic type measure of
(maximum – minimum)/2.
All QuickRead(s) do not return any message.
In case of error QuickRead abort the request.
All QuickRead(s) display a NAN value how digital number and the bargraph is off.

00DSEMI

Request from Host
Response from QuickRead(s)
Select Dynamic Type Measure of (Maximum – Minimum)/2.
All QuickRead(s) check the possibility and subsequently select dynamic type measure of
(maximum – minimum)/2.
All QuickRead(s) do not return any message.
In case of error QuickRead abort the request.
All QuickRead(s) display a NAN value how digital number and the bargraph is off.

00DCYCL

Request from Host
Response from QuickRead(s)
Start/Stop for Dynamic Cycle of Selected Measure.
All QuickRead(s) check the possibility and subsequently perform the start/stop for dynamic
cycle of selected measure.
All QuickRead(s) do not return any message.
In case of error QuickRead abort the request.
If is a start command all QuickRead(s) display the current value of dynamic measure
otherwise if is a stop command all QuickRead(s) are frozen on the last value of dynamic
measure.

00DSTOP

Request from Host
Response from QuickRead(s)
Stop for Any Dynamic Measure Cycle.
All QuickRead(s) check the possibility and subsequently perform the stop for any dynamic
measure cycle and leave dynamic mode.
All QuickRead(s) do not return any message.
In case of error QuickRead abort the request.
All QuickRead(s) display the value of static measure.
ASCII SERIAL TRANSMISSION PROTOCOL
00DMEAN
00DSEMI
00DCYCL
00DSTOP
201