Troubleshooting; Problems And Solutions - Welch Allyn Connex Vital Signs Monitor 6000 Serie Anexo A Las Instrucciones De Uso

Ocultar thumbs Ver también para Connex Vital Signs Monitor 6000 Serie:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 25
Search by clinician ID
Require password
Require clinician ID match to save measurements
5.
Do one of the following:

Troubleshooting

Patient data management messages
Message
Unable to identify patient.
Touch Clear to delete all data.
Unable to identify clinician.
Touch Clear to delete all data.

Problems and solutions

The problems addressed in this table do not generate alarm or information messages on the
monitor.
Problem
Save button is disabled
Start intervals button is
disabled
To continue in the Advanced Settings, touch another tab.
To exit the Advanced Settings and return to the Home tab, touch Exit.
Possible cause
The patient ID does not match an ID in the
patient list or network
The clinician ID does not match an ID in the
network
Possible cause
No patient ID
Patient search is in progress (the Patient
frame displays a process indicator)
The patient ID does not match an ID in the
patient list or network
Addendum to the directions for use, software versions 1.0X–1.7X 3
measurements from the Home tab. Clinician
information clears from the Clinician tab and the
Device Status area.
Note This setting does not take effect when
intervals are in progress.
Enable the monitor to query the network for
clinician information based on ID. The monitor
initiates the search when the clinician enters or
scans the ID from the Clinician tab. Returned
clinician information populates the Device Status
area and fields on the Clinician tab.
Require clinicians to enter their password, in
addition to ID, on the Clinician tab. The monitor
uses the ID and password combination to query the
network for clinician information.
Enable clinicians to save measurements only after
they have entered a clinician ID that matches an ID
in the network.
Suggested action
Re-enter the patient ID.
To delete all unsaved data,
Clear
touch
.
Re-enter the clinician ID.
To delete all unsaved data,
Clear
touch
.
Suggested action
Enter the patient ID.
Wait until the search is
complete.
Re-enter the patient ID.

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Connex integrated wall system

Tabla de contenido