Moria informs the availability of the One Use-Plus (OUP) handpiece for artificial chamber (#19175) used with One
Use Large-Cut heads (OU LC) (#19184/xxx) for single use or reusable artificial chamber.
1- Indications
OU LC heads for artificial chamber (#19184/xxx) guided for:
• RU (reusable) artificial chamber by the OUP guide ring for ALTK (#19173) mounted on ALTK artificial chamber
(#19161) equipped with its universal cover (#19162)
• SU (single-use) artificial chamber by the OUP LC single-use guide ring (#19187) mounted on single-use
artificial chamber (#19182)
enable to create, with an automated motion, a complete lamellar resection (without hinge) for donor graft
preparation during keratoplasty surgery.
2- Reference & Designation
• Reference: #19175
• Designation: OUP handpiece for artificial chamber
3- Instructions for use
1/ Prepare Evolution 3E control unit (see instruction for use #65090 included in box of OU LC heads, and user
manual Evolution 3E #65060/INTL):
• perform vacuum test on Evolution 3E control unit
• select OUP mode and speed 2
• connect OUP handpiece to Evolution 3E control unit (Fig. 1)
2/ See instruction for use #65090 for preparation of artificial chamber, setting of the graft and preparation
of the head.
3/ OUP handpiece setting:
• screw the head onto the handpiece (Fig. 2)
• inspect the blade under microscope and start the motor to check proper blade oscillation
• set OUP handpiece on OUP guide ring #19173 or #19187 depending on artificial chamber used (Fig. 3)
4/ See instruction for use #65090 for cutting of the graft and end of procedure.
4- Maintenance
For the maintenance of the OUP handpiece for artificial chamber (#19175), please refer to user manuals #65039
or #65040 (§9.3).
As for the cleaning-disinfection-sterilization-storage of OUP handpiece: see chart on following page.
Addendum for user manual #65009
& Instruction for use #65090
Tabla de contenido
loading

Resumen de contenidos para MORIA 19175

  • Página 5: Referencia Y Designación

    4/ Consulte las instrucciones de uso #65090 para cortar el injerto y finalizar el procedimiento. 4- Maintenance Para el mantenimiento de la pieza manual OUP para cámara artificial (#19175), por favor consulte los manuales del usuario #65039 o #65040 (§9.3).
  • Página 6: Avisos Importantes

    Pasos Qué hacer Imagen AVISOS IMPORTANTES Limpieza • La pieza manual debe limpiarse Fig.4 cuidadosamente después de cada • Para limpiar el tornillo del motor y procedimiento quirúrgico. las partes metálicas, utilice el cepillo desechable (#19149) humedecido con • No utilice abrasivos o espátulas para soluciones de limpieza.
  • Página 7 Fig.1 Fig.2 Fig.3 Fig.4      ...

Tabla de contenido