MANUAL DE INSTRUCCIONES CÓDIGO 80311 REV E
WATER
ACTIVATE
SETUP > PROBE
LANGUAGE
TIME/DATE
PRINTER
UNITS
PROBE
WATER
SETUP
Access to calibration and
configuration parameters.
LANGUAGE
TIME/DATE
PRINTER
UNITS
PROBE
WATER
SETUP
PROGRAMACIÓN
AJUSTES
PROGRAMMING
SETTINGS
TRABAJO
TASK
Cycle
number
PROGRAM O
PRO.: SOL. 2
TEMP.: 134
T.EST.: 4m
T.SEC.: 10m
SETUP
J
.P. SELECTA s.a.u.
8.2.6
With this parameter set in "activate" the autoclave, the level of the water tank
for steam generation will always be maintained in working limits. Manual filling
will not be necessary.
With the "disable" value set, the autoclave takes water for generating steam from
the autoclave tank itself. Filling additional deposit will be manual.
BACK
Press
HH:MM:SS DD/MM/YY
ter.
8.2.7
PASSWORD
The password protects the parameters that may affect the operation of the
autoclave.
CHANGE
To introduce the correct password, press over «INSERT PASSWORD».
PASSWORD
•
CAL
TEST
BACK
HH:MM:SS DD/MM/YY
8.3
The autoclave has two EN13060 test cycles programs to made easy operation
qualification:
•
PASSWORD
vacuum according to EN13060 (This test must be carried out with the cold boiler)
CHANGE
PASSWORD
•
into wrapped loads.
This test cycles should be run periodically, and results must be stored as quality
CAL
TEST
BACK
system records.
HH:MM:SS DD/MM/YY
Test
cycles
8.4
Up to 9 cycles could be stored on the autoclave, editable by the user.
The cycle sets the treatment to apply over samples, by means of the setting
parameters.
The 9 programs parameters are loaded by default in the device. See the
table in section 6.4 for the default values.
8.4.1
Cycle
- Select the cycle to modify by pressing over:
selection
icons
- The cycle (program) sets the sterilization treatment applied to the load, by
means of four parameters.
SEL
C
O
Note: Checking this box
the autoclave. Select another program and press the SEL key to select another
SAVE
of the programs in the list.
HH:MM:SS DD/MM/YY
Autovía A-2 Km 585.1 Abrera 08630 (Barcelona) España
e-mail: selecta@jpselecta.es - website: http://www.jpselecta.es
Ene 2020
(Sujetas a modificaciones sin previo aviso)
Activates / disables external water entry
to return to the main screen or
Parameters protected by a password
Calibration of measuring devices (password "0000"):
Internal temp: offset value (Offset) of the water probe
External Temp: offset value (Offset) of the heart probe
Pressure: displacement value (Offset) of the pressure probe
Pre-heat: preheating temperature of the boiler
Program 0-8
Temperature (default 95ºC)
Log interval: time between the registration of two parameters 1-60min
Test cycles (Vacuum test & Bowie Dick)
Vacuum test: Qualifies the autoclave tightness and its capability to get suitable
BOWIE DICK test: Qualifies the steam quality and steam capability to penetrate
Sterilization programs
Modify & save a program
-
Purge type.
-
Sterilization temperature.
-
Sterilization time.
-
Drying time.
SEL
to modify another parame-
BACK
which is the operating program when starting
Tel 34 93 770 08 77 Fax 34 93 770 23 62
Pag.: 35