R&S ESH2-Z3

3 Required Measuring Equipment

Item
Designation, required specifications
Recommended R&S equipment
1
Network analyzer 9 kHz..100 MHz
2
Calibration kit
3
2 x BNC measurement cable
4
2 x BNC connector
5
BNC-T connector
6
Adapter BNC-banana plug

Recalibration

Recalibration of the Passive Voltage Probe R&S ESH2-Z3 is carried out by a two-port measurement
using the test setup shown below.
Test setup:
Remove crocodile clamp from earth cable. So the banana plug can be put into the BNC-banana
adapter.
ZVR
BNC-T
Adapter BNC-banana plug
Cable1
Measurement:
)
feed through (S21
9 kHz..30 MHz:.................................................................................. -30 dB +1/-5 dB
Typical result of S21:
1156.6124.32
Type
R&S ZVR
Cable2
BNC connector
50 Ohm
red
ESH2-Z3 with
black
main voltage tip
Required Measuring Equipment
Order No.
1127.8551.61
0017.6559.00
0017.6588.00
0017.6742.00