* DRY FUNCTION NOT PERFECT:
Please open the door immediately, in 15 minutes, after alarm. If still have same problem, please call
engineer.
PS: ANY PROBLEMS PLEASE CHECK THE WIRE CONNECTION AT FIRST.
CHAPTER 8 MAINTENANCE REQUIREMENTS
DAILY:
- Wipe the inside of the chamber, door and the gasket with a damp, lint-free cloth.
- Check the water level. Top up with distilled water only.
WEEKLY:
- Drain off the remaining water of water reservoir and refill with distilled water. (*Please use the
water level viewing hose to drain off water of water reservoir.)
- Remove the cap of water reservoir and find the safety valve. Pull the ringlet of the safety valve to
check its working status.
MONTHLY:
- Clean the chamber and piping system with "CHAM-MATE".
- Use clean water to scrub the water reservoir thoroughly. (It's to clean away the mineral residues of
water.)
- The water filter cleaning gate at rear of the machine should be cleaned.
ANNUALLY:
(Please contact your distributor for maintenance.)
- Check the working status of silicone door gasket, steam trap, safety valve, and heater. (Recommen-
dation: Replace the silicone door gasket every year.)
- Test the functions of indicator, solenoid valve.
- Check the temperature during sterilization.
- Use biological indicators to test the validity of sterilization.
WARRANTY
Quirumed offers the guarantee provided by law against all manufacturing defects from the date of
purchase. You should therefore retain the invoice or purchase.
When using non-original parts or modifies the product immediately void.
Misuse or wear of product is not considered a manufacturing defect and hence not included in the
warranty.
Nor a bad unpacking the product included such cuts made with cutter or casualties.
For any matter relating to the warranty please contact your dealer.
16
www.quirumed.com